  "recall_number": "Z-0595-2026",
  "product_description": "FlexLab (FLX);  Version: FLX-217-10;",
  "company_name": "Inpeco S.A.",
  "reason": "The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results.  This error occurs if the operators open the printer head of  mALX to perform error recoveries or maintenance procedures and do not empty the printer  tubes buffer to remove all the secondary sample tubes.",
  "classification": "Class II",
  "status": "Ongoing",
  "source": "Devices",
  "initiated_date": "2025-10-30T00:00:00.000Z",
  "slug": "flexlab-flx-version-flx-217-10-z-0595-2026",