  "recall_number": "Z-0775-2023",
  "product_description": "Programmable Diagnostic Computer",
  "company_name": "Siemens Medical Solutions USA, Inc",
  "reason": "The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues:     1) \"PASSWORD STORE CORRUPTED\" error message during system boot;     2) Subsystem crash during examination;    3) Dialog Monitor Computer (DMC) application crash while loading a study; and    4) Software crash due to system internal timeout.     Issue 1 may lead to a delay or interruption of procedure. Issues 2, 3, and 4 may result in delay in starting or continuing the examination, and may also prevent the operator from starting or continuing a study;",
  "classification": "Class II",
  "status": "Ongoing",
  "source": "Devices",
  "initiated_date": "2022-11-16T00:00:00.000Z",
  "slug": "programmable-diagnostic-computer-z-0775-2023",
  "matched_lei": "549300IVVSJMM2J81G56",
  "updated_at": "2026-07-16T13:49:11.326Z"
}